GDD-1A Automatic Oil Moisture Content Analyzer According to Karl Fischer Method


GDD-1A Automatic Oil Moisture Content Analyzer According to Karl Fischer Method

Introduction:

  • The instrument  Automatic Titrator for Laboratory / Laboratory Automatic Titrator  adopts the newest mechanical-electronic technology with humanistic interface. 


  • It is reliable with wide measuring range(0.003%-100%), high precision and affordable for users. 


  • It can determine crystal water, adsorbent water in gas, liquid, solid samples. 


  • It can automatically analyze, absorb and discharge liquid, adjust instrument failure, protect important parts, calibrate electrode parameters, etc. 

  • GDD-1A Automatic Oil Moisture Content Analyzer According to Karl Fischer Method  

    Features:

  • It has dual platinum electrode polarization voltage to judge and show the reaction endpoint,then stop titration

  • Fully sealed system, in order to avoid ambient moisture's penetration.

  • By the function keys,the instrument can absorb reagent, measure, show end point, absorb solvent, discharge waste fluid, stir and adjust speed. 

  • Displaying the mixed results simultaneously, titration dynamically display and automatic memory of test data.

  • Automatically deduct drifting value,automatically track the ambient moisture drifting and make sure the test data is precise.

  • The system is fully sealed,it can renew the solvent and discharge waste fluid  automatically, to avoid leak out toxic solvent.

  • Instrument status, dip stool sample volume, present time and drifting value,etc could be displayed instantly.

  • The instrument runs smoothly, low noise

  •  

    GDD-1A Automatic Oil Moisture Content Analyzer According to Karl Fischer Method   

    Specifications:

     

    Display

    Large LCD screen

    Testing method

    Volumetric method Karl Fischer

    Measurement range

    0.003%-100%(water phase content)

    Resolution

    1ug

    Repeatability

    <0.01

    Titration linear coefficient

    ≥0.998

    Test time

    2~5 min

    End-point indicate

    Screen display, print out, alarm,end-point indicator light

    Printer

    Thermal printer

    Data storage

    100 groups

    Burette

    200ml automatic burette

    End-point Delay Time

    10s-300s adjustable continuously

    Titration accuracy

    0.001ml

    Working environment

    5~40 Centigrade, humidity<85%

    Power supply

    AC220V+/-10%, 50Hz+/-2.5Hz

    Gross weight

    5.5kg


bigPhoto
 Mr. Boswell

Tel: 86-23-98954970
Contact to this supplier

Related Search

Find more related products in following catalogs on Hisupplier.com

Related Products

Company Info

Chongqing Gold M&E Equipment Co.,Ltd [China (Mainland)]


Business Type:Manufacturer, Trading Company
City: Chongqing
Province/State: Chongqing
Country/Region: China (Mainland)

You May Like:

Product (150)